Dansk

TTS 500FA

Replaces all previous types: LTT420M, STT1000SA, TS334 and TS334S

Clamp-on temperature measurement.

Quick configuration with simple windows software by
connecting the PC´s USB port directly to the device´s M12 connector (4 pole).

Technical data:

Supply voltage : 7 - 34VDC

Measuring range °C : -40 til 180 °C ( Configuration resolution = 0,1°C (IEC 751)

Output in transmitter-mode : I - Load = 4-20 mA (effect of variations in supply voltage≤ 0,1µA/V)

Output in  switch-mode : I - Load off = 3,40 mA / I load on = 34,00 mA ± 1%

Ambient temperature coefficient  : ≤ ± 0.002 % af  F.S. / °C (from -30 to +80 °C)

Direct invertet function : Optional     

Hysteresis from switch on to off : Free configurable by 0,1°C resolution. (symmetrical around  set point´s value, minimum value  = 0,2°C)

Temperature sensor : PT100 1/3 din B curve (accuracy in °C ≤ ± ( 0,1 (measured temperature x 0,0017))  

Accuracy on electronic unit : Better than ± 0,1°C according to the IEC751 standard.

Calibration facilities : With PC connected, Off set and Gain can be adjusted. ( with Off set = 0,00 °C and Gain = 1,0000 the IEC 751 standard is followed).

2 color LED indicator (green/red) : LED in side of the house indicate the device phases during operation.

Physical dimentions mm : B x H x L = 22 x 30 x 38

Download PC application TTS500A (zip)

Download PC application TTS500A (exe)


Products

Did you know?

That Sensonic temperature gauges can measure temperature in a process pipe without welding a pocket?